INSTRUCTION BOOK
Terminating Power Sensors
&
Extended Frequency
Terminating Power Sensors
Models 5011, 5011-EF,
5015 and 5015-EF
©Copyright 2009 by Bird Electronic Corporation
Instruction Book Part Number 920-5011S Rev. A
Zobrazit stránku 0
1 2 3 4 5 6 ... 30 31

Shrnutí obsahu

Strany 1 - Extended Frequency

INSTRUCTION BOOKTerminating Power Sensors&Extended FrequencyTerminating Power SensorsModels 5011, 5011-EF,5015 and 5015-EF©Copyright 2009 by Bird

Strany 3 - Safety Symbols

ixTable of ContentsSafety Precautions 0Safety Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . iWarning Statements . . . . . . . . .

Strany 5 - Safety Statements

1Chapter 1 IntroductionDescriptionThe Bird 5015 Terminating Power Sensor (TPS) is a diode-based power sensor that measures true average power from 40

Strany 7

3Chapter 2 InstallationConnecting Terminating Power Sensor (TPS) Note: Connections are the same for the Bird 5011 and 5011-EF. CAUTIONDischarge all

Strany 8

41. Do one of the following:z To connect to a Digital Power Meter:z For Models 5011 and 5011EF: Connect the Bird TPS to the “Sensor” serial port on t

Strany 9 - Literature Contents

5Chapter 3 OperationZeroing SensorOver time, the sensor’s “zero value” (reading with no applied RF power) can drift, making all readings inac-curate b

Strany 10

6Example - If a sensor has a specified accuracy of 5% of reading + 1.0 uW, then for a 10 mW signal the uncertainty is ± 0.501 mW. For a 1 mW signal th

Strany 11 - Table of Contents

7Table 2 - External FactorsError Source ConditionsThe root sum square (RSS) uncertainty is the industry standard method for combining independent unce

Strany 12

0Safety PrecautionsThe following are general safety precautions that are not necessarily related to any specific part or procedure, and do not necessa

Strany 13 - Chapter 1 Introduction

8Table 3 has two examples of uncertainty calculations. The first is a 10 mW signal at room temperature. The second is a 10 µW, 40 MHz signal at 50°C.

Strany 14

9Mismatch UncertaintyAnother factor of measurement accuracy is mismatch uncertainty. When a source and a load have different impedances, some signal w

Strany 15 - Chapter 2 Installation

10 Note: If a source with a 1.30:1 VSWR was used instead, the mismatch uncertainty would be:Using a lower VSWR source can drastically reduce the mi

Strany 16

11Chapter 4 Specifications5011 SpecificationsPower Measurement CharacteristicsGeneral Terminated average powerFrequency Range 40 MHz to 4 GHzPower Mea

Strany 17 - Chapter 3 Operation

125011-EF SpecificationsPhysical and Environmental SpecificationsOperating Temp. –10 to +50 °C (+14 to +122 °F)Storage Temp. –40 to +80 °C (–40 to +17

Strany 18

13Input Impedance 50 Ohms (nominal)Input VSWR: Typical Maximum 1.05 (32.0 dB return loss) 1.25 (19.1 dB return loss)Input Connector Precision N MaleOu

Strany 19

14®Series 5011™Terminating Power SensorBird Elec tro nic Corp., Solon , Ohio USAWeb: http://www.bird-electronic.com4 1/4"[108 mm]6”[152 mm]1 3/8”

Strany 20

155015 Specifications Power Measurement CharacteristicsGeneral Terminated average powerFrequency Range 40 MHz to 4 GHzPower Measurement Range–20.0 to

Strany 21 -  Note: where ρ

165015-EF SpecificationsPhysical and Environmental SpecificationsOperating Temp. –10 to +50 °C (+14 to +122 °F)Storage Temp. –40 to +80 °C (–40 to +17

Strany 22

Physical and Environmental Specifications17Input VSWR: Typical Maximum 1.05 (32.0 dB return loss) 1.25 (19.1 dB return loss)Input Connector Precision

Strany 23 - Chapter 4 Specifications

iSafety Symbols Note: Calls attention to supplemental information.WARNINGSafety Earth GroundAn uniterruptible earth safety ground must be supplied

Strany 25

19Limited WarrantyAll products manufactured by Seller are warranted to be free from defects in material and workmanship for a period of one (1) year,

Strany 26 - Terminating Power Sensor

iiWarning StatementsThe following safety warnings appear in the text where there is danger to operating and maintenance personnel, and are repeated he

Strany 27 - 5015 Specifications

iiiSafety StatementsUSAGEANY USE OF THIS INSTRUMENT IN A MANNER NOT SPECIFIED BY THE MANUFACTURER MAY IMPAIR THE INSTRUMENT’S SAFETY PROTECTION.USOEL

Strany 28 - 5015-EF Specifications

ivSERVICESERVICING INSTRUCTIONS ARE FOR USE BY SERVICE - TRAINED PERSONNEL ONLY. TO AVOID DANGEROUS ELECTRIC SHOCK, DO NOT PERFORM ANY SERVICING UNLES

Strany 29

vASSISTENZA TECNICALE ISTRUZIONI RELATIVE ALL’ASSISTENZA SONO PREVISTE ESCLUSIVAMENTE PER IL PERSONALE OPPORTUNAMENTE ADDESTRATO. PER EVITARE PERICOLO

Strany 30

viRF VOLTAGE MAY BE PRESENT IN RF ELEMENT SOCKET - KEEP ELEMENT IN SOCKET DURING OPERATION.DE LA TENSION H.F. PEAT ÊTRE PRÉSENTE DANS LA PRISE DE L&ap

Strany 31 - Limited Warranty

viiAbout This ManualThis manual covers the operating and maintenance instructions for the following models:Changes to this ManualWe have made every ef